Investigating why vacuum was weak at the tip support.
DD engineers are experienced in getting to the route of the problem.
#jasonsagarengineer#DDengineer https://t.co/liPH3z2wtD
Adec Performer chair with intermittent suction
Problem traced to the air electric switch ,
Problem fixed and surgery now available for use in this extremely busy week
# #dentalengineer#adecperformer#adecuk#DDengineer https://t.co/ZlmV4AF8Q4
As with all visits which relate to suction issues, we ensure our customers are achieving adequate vacuum before leaving site.
DD engineers are fully competent to do this and resolve any issues promptly.
#jasoncsagar#DDengineer#dentalsuctiontesting#dent…https://t.co/KV24ggAH0k
Kavo E50 fault today, two issues found , Aquamat container bone dry and thick with biofilm blocking the inlet.
Plus the solenoid had split creating a slow drip.
#kavokerr#cardiffdentalengineer https://t.co/8DO0md7X2l
Installation in Birmingham today
Belmont 097 Belray II
The wall was not strong enough to use rawlbolts, so the use of threaded 10mm bar and Chemfix was enough to provide a firm strong fixing of the wall plate.
#jasonsagarengineer#dentalengineer#cardi…https://t.co/mbcLOWGfV0
Hi Jason
Thank you for coming last week it really was appreciated.
This appreciation was received today from the PM at Village Dental Practice in Wales.
There suction was down and after some investigation on site found the issue to lie with the chair.
#…https://t.co/jQ3TOoZ7cR