[Notice] AV500 is updated - topspin 4.4.0 on Almalinux9
Please contact NMR manager for any inconvenience after update.
AV500장비를 업데이트하였습니다
(Topspin 4.4.0 on Almalinux9)
사용에 불편이 있으시면 NMR관리자에게 연락하여 주십시오.
[Notice] AVNEO400 is updated
- topspin 4.4.0 on Almalinux9
- probe is cleaned
Please contact NMR manager for any inconvenience after update.
AVNEO400장비를 업데이트하였습니다 (Topspin 4.40 on Almalinux9 & probe청소)
사용에 불편이 있으시면 NMR관리자에게 연락하여 주십시오.
[2023.06.14 11:00/IBS600] IBS600 is running after maintenance.
- broken NMR tube is removed.
- upper barrel and probe are cleaned
- shimmap, lock frequency, standard shim are updated
[2023.06.13 19:00/IBS600] Sample is broken in the magnet. Reservation is blocked to prevent further accident. IBS600 will be re-running after maintenance tomorrow morning.
@pedromiguel13 Yes, I agree that mixing of spinner is not perfectly correlated. Sensing problem may happen when sensor is damaged or dust blocks spin air line of probe, spinner label is erased. In my case, trouble happens just after autosampler installation so I guess old spinner is main cause
[AS400] Do not bring spinner of other NMR to AS400. Icon-NMR can be stopped with old spinner(recognition trouble) Use only spinners of AS400 spinner tray.
@pedromiguel13 Old spinner trouble happens on Samplecase plus(60 tray) frequently. Also
Samplecase(24 tray) with old firmware(when 2018~2019) have no trouble with old spinner, but with new firmware(when 2022) have same error sometimes.
To prevent this trouble, old spinners (yellow color of label is more darker than new one) were temporarily removed from all NMR spinner tray until changing label.
[2023.05.02 19:00/AVHD400] Some user put NMR tube into magnet without spinner.
Reservation and workstation of AVHD400 are closed to prevent further accident. AVHD400 will be available after maintenance is finished (expectation : 2023-05-03 10:00)