@johnhammockmd You really need a good rate response algorithm. CLS works the best for VVS. Resting rate needs to be deactivated. Anecdotally I’ve seen it works well. Spain study. https://t.co/4iz3jGgm9H
@PeterJesselEP 17 sub threshold pulses. It’s Biotronik rate response algorithm CLS at work. You can filter these out on your ECG machine. https://t.co/h4XrC8E9f3
@MelbourneEPdoc Can’t resterilise. We do get gifted some out of date SR PPM’s for this purpose. They remain in the expired package until needed and discarded after use. Regarded as unsterile and placed on top of the large op site dressing that covers external portion of the lead.
2 previous failed ablation attempts in other centers. Repetitive non-sustain VT episodes. On Holter, > 85% of beats were related VT. Sinus beat is conducted with LBBB(fig1). During mapping, transient AVB due to catheter bump in RV(Fig2) Activation map and earliest site.